Markus Klotz GmbH
Theodor-Heuss-Straße 27
75378 Bad Liebenzell
fon +49(0)7052 / 9 23 36
fax +49(0)7052 / 9 23 38
info@fa-klotz.de
www.fa-klotz.de
Spec. sheet "particle counting system PZG-4"
download here
The sample feeder PZG 4 with the PC insert board PCI and the evaluation
software SW-OI and the double piston pump system (DPS) are employed for
measuring oils and liquids from containers. The representation and
storage of measured data is carried out on the PC according to the
selected standard.
By using the pressure container (pressure dome) with the special Klotz
sealing system sample bottles can be very simply degassed by applying
vacuum and can then be fed, pressurized, through the sensor. The DPS
ensures the accurate feed of the medium.
For oils with viscosities of up to 350 mm2/s the LDS 45/50 laser sensor (particle concentrations max. 60.000 particles/ml [>=4μm] is employed. For low viscosity liquids measurements can be carried out alternatively with the LDS 30/30 sensor particle concentrations max. 130.000 particles/ml [>=4μm]; resolutions <5%).
The SW-OI program allows you to carry out measurements with the sample
feeder PZG4. It was specifically designed for measurements of oils. It
can carry out evaluations according to ISO 4406, ISO (NIST) and NAS.
Average determination, protocol printouts as well as tabular
representations are possible. The measuring and flushing volumes can be
given in multitudes of 10ml since with every pump stroke exactly 10ml
liquid is fed.
A flexible export to Excel for further evaluations and representations is also possible.
Special pump for small volumes and USB interface
Contamination control of hydraulic oils, transmission oils, diesel, washing liquids
Markus Klotz GmbH . Theodor-Heuss-Straße 27 . 75378 Bad Liebenzell
fon +497052-92336 . fax +497052-92338 . info@fa-klotz.de